Variation of system performance, quality control standards and adherence to international FDG-PET/CT imaging guidelines. A national survey of PET/CT operations in Austria

Nuklearmedizin. 2014;53(6):242-8. doi: 10.3413/Nukmed-0665-14-05. Epub 2014 Aug 18.

Abstract

Aim: To gather information on clinical operations, quality control (QC) standards and adoption of guidelines for FDG-PET/CT imaging in Austrian PET/CT centres.

Methods: A written survey composed of 68 questions related to A) PET/CT centre and installation, B) standard protocol parameters for FDG-PET/CT imaging of oncology patients, and C) standard QC procedures was conducted between November and December 2013 among all Austrian PET/CT centres. In addition, a NEMA-NU2 2012 image quality phantom test was performed using standard whole-body imaging settings on all PET/CT systems with a lesion-to-background ratio of 4. Recovery coefficients (RC) were calculated for each lesion and PET/CT system.

Results: A) 13 PET/CT systems were installed in 12 nuclear medicine departments at public hospitals. B) Average fasting prior to FDG-PET/CT was 7.6 (4-12) h. All sites measured blood glucose levels while using different cut-off levels (64%: 150 mg/dl). Weight-based activity injection was performed at 83% sites with a mean FDG activity of 4.1 MBq/kg. Average FDG uptake time was 55 (45-75) min. All sites employed CT contrast agents (variation from 1%-95% of the patients). All sites reported SUV-max. C) Frequency of QC tests varied significantly and QC phantom measurements revealed significant differences in RCs.

Conclusion: Significant variations in FDG-PET/CT protocol parameters among all Austrian PET/CT users were observed. Subsequently, efforts need to be put in place to further standardize imaging protocols. At a minimum clinical PET/CT operations should ensure compliance with existing guidelines. Further, standardized QC procedures must be followed to improve quantitative accuracy across PET/CT centres.

Keywords: Austria; FDG; PET/CT; multicentre; protocols; quality control.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Austria
  • Fluorodeoxyglucose F18*
  • Guideline Adherence / statistics & numerical data*
  • Health Care Surveys
  • Internationality
  • Multimodal Imaging / standards*
  • Multimodal Imaging / statistics & numerical data
  • Nuclear Medicine / standards*
  • Positron-Emission Tomography / standards*
  • Positron-Emission Tomography / statistics & numerical data
  • Practice Guidelines as Topic
  • Radiopharmaceuticals
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / standards*
  • Tomography, X-Ray Computed / statistics & numerical data

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18