Application of an electrochemical immunosensor with a MWCNT/PDAA modified electrode for detection of serum trypsin

Sensors (Basel). 2014 Jun 10;14(6):10203-12. doi: 10.3390/s140610203.

Abstract

Objective: To establish an electrochemical immunosensor for the determination of serum trypsin levels using a multiwall carbon nanotubes (MWCNTs)-composite-modified electrode.

Method: A MWCNT composite coated on the surface of bare gold electrodes was used for fixation of an anti-trypsin antibody. The assembly process and the performance indicators, including sensitivity, linear range of detection, anti-jamming performance, and stability, of the electrochemical immunosensor were examined by cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS).

Results: With optimized experimental conditions, the difference of the current value measured by differential pulse voltammetry (DPV) showed a linear relationship with the concentration of serum trypsin within 0.10-100 ng/mL. The detection limit for trypsin using this sensor was 0.002 ng/mL.

Conclusions: The electrochemical immunosensor built using the MWCNT-composite-modified electrode is simple to operate and has a fast response time, along with a wide linear range, high sensitivity, and accuracy, making it suitable for serum trypsin detection.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrochemical Techniques / instrumentation*
  • Electrochemical Techniques / methods
  • Electrodes
  • Gold
  • Humans
  • Immunoassay / instrumentation*
  • Immunoassay / methods
  • Nanotubes, Carbon / chemistry*
  • Polyethylenes
  • Quaternary Ammonium Compounds
  • Trypsin / blood*

Substances

  • Nanotubes, Carbon
  • Polyethylenes
  • Quaternary Ammonium Compounds
  • poly-N,N-dimethyl-N,N-diallylammonium chloride
  • Gold
  • Trypsin