Preclinical study of a biodegradable polymer-based stent with abluminal sirolimus release

Arq Bras Cardiol. 2014 May;102(5):432-40. doi: 10.5935/abc.20140044. Epub 2014 Apr 17.
[Article in English, Portuguese]

Abstract

Background: Bioabsorbable polymer stents with drug elution only on the abluminal surface may be safer than durable polymer drug-eluting stents.

Objective: To report the experimental findings with the Inspiron™ stent - a bioabsorbable polymer-coated stent with sirolimus release from the abluminal surface only, recently approved for clinical use.

Methods: 45 stents were implanted in the coronary arteries of 15 pigs. On day 28 after implantation, angiographic, intracoronary ultrasonographic and histomorphological data were collected. Five groups were analyzed: Group I (nine bare-metal stents); Group II (nine coated with bioabsorbable polymer on the luminal and abluminal surfaces); Group III (eight stents coated with bioabsorbable polymer on the abluminal surface); Group IV (nine stents with bioabsorbable polymer and sirolimus on the luminal and abluminal surfaces); and Group V (ten stents with bioabsorbable polymer and sirolimus only on the abluminal surface).

Results: The following results were observed for Groups I, II, III, IV and V, respectively: percentage stenosis of 29 ± 20; 36 ± 14; 33 ± 19; 22 ± 13 and 26 ± 15 (p = 0.443); late lumen loss (in mm) of 1.02 ± 0.60; 1.24 ± 0.48; 1.11 ± 0.54; 0.72 ± 0.44 and 0.78 ± 0.39 (p = 0.253); neointimal area (in mm(2)) of 2.60 ± 1.99; 2.74 ± 1.51; 2.74 ± 1.30; 1.30 ± 1.14 and 0.97 ± 0.84 (p = 0.001; Groups IV and V versus Groups I, II and III); and percentage neointimal area of 35 ± 25; 38 ± 18; 39 ± 19; 19 ± 18 and 15 ± 12 (p = 0.001; Groups IV and V versus Groups I, II and III). Injury and inflammation scores were low and with no differences between the groups.

Conclusion: The Inspiron™ stent proved to be safe and was able to significantly inhibit the neointimal hyperplasia observed on day 28 after implantation in porcine coronary arteries.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorbable Implants*
  • Animals
  • Biopolymers*
  • Coated Materials, Biocompatible*
  • Coronary Angiography
  • Coronary Vessels / diagnostic imaging
  • Coronary Vessels / drug effects*
  • Drug Delivery Systems / methods
  • Drug-Eluting Stents*
  • Immunosuppressive Agents / administration & dosage*
  • Materials Testing
  • Reference Values
  • Reproducibility of Results
  • Sirolimus / administration & dosage*
  • Swine
  • Time Factors
  • Ultrasonography

Substances

  • Biopolymers
  • Coated Materials, Biocompatible
  • Immunosuppressive Agents
  • Sirolimus