Novel inductive decoupling technique for flexible transceiver arrays of monolithic transmission line resonators

Magn Reson Med. 2015 Apr;73(4):1669-81. doi: 10.1002/mrm.25260. Epub 2014 Apr 17.

Abstract

Purpose: This article presents a novel inductive decoupling technique for form-fitting coil arrays of monolithic transmission line resonators, which target biomedical applications requiring high signal-to-noise ratio over a large field of view to image anatomical structures varying in size and shape from patient to patient.

Methods: Individual transmission line resonator elements are mutually decoupled using magnetic flux sharing by overlapping annexes. This decoupling technique was evaluated by electromagnetic simulations and bench measurements for two- and four-element arrays, comparing single- and double-gap transmission line resonator designs, combined either with a basic capacitive matching scheme or inductive pickup loop matching. The best performing array was used in 7T MRI experiments demonstrating its form-fitting ability and parallel imaging potential.

Results: The inductively matched double-gap transmission line resonator array provided the best decoupling efficiency in simulations and bench measurements (<-15 dB). The decoupling and parallel imaging performance proved robust against mechanical deformation of the array.

Conclusion: The presented decoupling technique combines the robustness of conventional overlap decoupling regarding coil loading and operating frequency with the extended field of view of nonoverlapped coils. While demonstrated on four-element arrays, it can be easily expanded to fabricate readily decoupled form-fitting 2D arrays with an arbitrary number of elements in a single etching process.

Keywords: coil array; mutual decoupling; pickup loop matching; surface coil; transmission line resonator; ultrahigh field.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electronics, Medical / instrumentation*
  • Equipment Design
  • Equipment Failure Analysis
  • Image Enhancement / instrumentation*
  • Magnetic Resonance Imaging / instrumentation*
  • Magnetics / instrumentation*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Transducers*