A simplified echocardiographic technique for detecting continuous-flow left ventricular assist device malfunction due to pump thrombosis

J Heart Lung Transplant. 2014 Jun;33(6):575-86. doi: 10.1016/j.healun.2014.01.865. Epub 2014 Jan 28.

Abstract

Background: Malfunction of a continuous-flow left ventricular assist device (CF-LVAD) due to device thrombosis is a potentially life-threatening event that currently presents a diagnostic challenge. We aimed to propose a practical echocardiographic assessment to diagnose LVAD malfunction secondary to pump thrombosis.

Methods: Among 52 patients implanted with a CF-LVAD from a single center who underwent echocardiographic pump speed-change testing, 12 had suspected pump thrombosis as determined by clinical, laboratory, and/or device parameters. Comprehensive echocardiographic evaluation was performed at baseline pump speed and at each 1,000-rpm interval from the low setting of 8,000 rpm to the high setting of 11,000 rpm in 11 of these patients.

Results: Receiver operating characteristic curves and stepwise logistic regression analyses showed that the best diagnostic parameters included changes in the LV end-diastolic diameter (<0.6 cm), aortic valve opening time (<80 msec), and deceleration time of mitral inflow (<70 msec) from lowest to highest pump speed. One parameter was predictive of pump malfunction, with 100% sensitivity and 89% specificity, whereas 2 of 3 parameters increased the sensitivity to 100% and specificity to 95%.

Conclusions: The 3 echocardiographic variables of measured changes in LV end-diastolic diameter, aortic valve opening time, and deceleration time of mitral inflow between the lowest (8,000 rpm) and highest pump speed settings (11,000 rpm) during echo-guided pump speed-change testing appear highly accurate in diagnosing device malfunction in the setting of pump thrombosis among patients supported with CF-LVAD. Further investigation is warranted to create and validate a prediction score.

Keywords: cardiomyopathy; echocardiography; heart failure; left ventricular assist device; thrombosis.

MeSH terms

  • Adult
  • Aged
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Blood Flow Velocity / physiology
  • Echocardiography / methods*
  • Equipment Failure*
  • Female
  • Heart Failure / diagnostic imaging
  • Heart Failure / therapy*
  • Heart-Assist Devices / adverse effects*
  • Humans
  • Logistic Models
  • Male
  • Middle Aged
  • Mitral Valve / diagnostic imaging
  • Mitral Valve / physiopathology
  • Predictive Value of Tests
  • ROC Curve
  • Retrospective Studies
  • Thrombosis / diagnosis*
  • Thrombosis / physiopathology
  • Ventricular Function, Left / physiology