Validation of Serological Testing for Anti-Treponema pallidum from Postmortem Blood on the Siemens-BEP(®)-III Automatic System

Transfus Med Hemother. 2013 Dec;40(6):403-8. doi: 10.1159/000354070. Epub 2013 Oct 17.

Abstract

Background: Infectious disease marker testing is obligatory for the release of human tissue for transplantation. Most CE-marked tests are not validated for postmortem blood. In a previous study we have validated the testing for anti-HIV-1/2, anti-HCV, HBsAg, and anti-HBc. Here, we present the validation of testing for antibodies against T. pallidum, which is the last marker obligatory for tissue release for transplantation.

Methods: 17 samples of postmortem sera and 10 samples of both pre-und postmortem sera were obtained from cornea donors and tested for anti-T. pallidum on the Siemens-BEP-III-System. These sera were spiked with anti-T. pallidum-positive standard sera in concentrations which give low- and high-positive results at the respective dilution.

Results: Two of the unspiked postmortem sera were false-positive most likely due to intense hemolysis (free hemoglobin > 50 mg/dl). Of the 25 negative postmortem sera, none of the spiked samples was false-negative after 0, 24 and 60 h.

Conclusion: There is no indication that postmortem samples give false-negative or false-positive results with the test system and test kits used in cases of low hemolysis. The procedure described might serve as a model for validating other test kits on postmortem samples.

Keywords: Blood testing; Infectious disease serology; Postmortem blood; Tissue donation; Treponema pallidum.