Filtering and thresholding the analytic signal envelope in order to improve peak and spike noise reduction in EEG signals

Med Eng Phys. 2014 Apr;36(4):547-53. doi: 10.1016/j.medengphy.2013.11.014. Epub 2013 Dec 21.

Abstract

To remove peak and spike artifacts in biological time series has represented a hard challenge in the last decades. Several methods have been implemented mainly based on adaptive filtering in order to solve this problem. This work presents an algorithm for removing peak and spike artifacts based on a threshold built on the analytic signal envelope. The algorithm was tested on simulated and real EEG signals that contain peak and spike artifacts with random amplitude and frequency occurrence. The performance of the filter was compared with commonly used adaptive filters. Three indexes were used for testing the performance of the filters: Correlation coefficient (ρ), mean of coherence function (C), and rate of absolute error (RAE). All these indexes were calculated between filtered signal and original signal without noise. It was found that the new proposed filter was able to reduce the amplitude of peak and spike artifacts with ρ>0.85, C>0.8, and RAE<0.5. These values were significantly better than the performance of LMS adaptive filter (ρ<0.85, C<0.6, and RAE>1).

Keywords: Biomedical signal processing; Digital filters; Electroencephalography.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Artifacts*
  • Brain / physiology
  • Computer Simulation
  • Electroencephalography / methods*
  • Humans
  • Signal Processing, Computer-Assisted*