Sensitivity and accuracy of organic matrix-assisted laser desorption and ionisation mass spectrometry of FeCl3 is higher than in in matrix-free approach

Eur J Mass Spectrom (Chichester). 2013;19(2):77-89. doi: 10.1255/ejms.1217.

Abstract

We compare the quality and reliability of laser desorption and ionisation mass spectra of FeCl3 acquired without the assistance of the matrix with the spectra acquired with different organic matrix molecules. Generally, inorganic salts tend to form clusters upon laser irradiation, the signals of which can be easily distinguished from ions arising from the matrix. In the presence of a matrix, cluster ions are, however, mostly suppressed. We have compared the number of analyte signals, accuracy of determination of isotope composition of the analyte and the sensitivity of FeCl3 detection between different approaches. The results obtained imply that the sensitivity of mass spectrometric analysis of FeCl3 is somewhat higher when matrices are applied than in the matrix-free approach. Among all matrices tested in this work, F20TPP seems to be the most promising for further applications as a matrix for mass spectrometry of inorganic salts.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chlorides / analysis*
  • Ferric Compounds / analysis*
  • Indicators and Reagents
  • Iron Radioisotopes / chemistry
  • Limit of Detection
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / methods*