Effects of RF profile on precision of quantitative T2 mapping using dual-echo steady-state acquisition

Magn Reson Imaging. 2014 Jan;32(1):102-6. doi: 10.1016/j.mri.2013.10.002. Epub 2013 Oct 10.

Abstract

The dual echo steady-state (DESS) sequence has been shown successful in achieving fast T2 mapping with good precision. Under-estimation of T2, however, becomes increasingly prominent as the flip angle decreases. In 3D DESS imaging, therefore, the derived T2 values would become a function of the slice location in the presence of non-ideal slice profile of the excitation RF pulse. Furthermore, the pattern of slice-dependent variation in T2 estimates is dependent on the RF pulse waveform. Multi-slice 2D DESS imaging provides better inter-slice consistency, but the signal intensity is subject to integrated effects of within-slice distribution of the actual flip angle. Consequently, T2 measured using 2D DESS is prone to inaccuracy even at the designated flip angle of 90°. In this study, both phantom and human experiments demonstrate the above phenomena in good agreement with model prediction.

Keywords: 2D and 3D imaging; Dual-echo steady-state imaging; Flip angle dependency; Slice profile; T2 quantification.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brain / pathology
  • Computer Simulation
  • Humans
  • Image Processing, Computer-Assisted
  • Imaging, Three-Dimensional / methods
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Reproducibility of Results