Innovative application of available stent grafts in Japan in aortic aneurysm treatment-significance of innovative debranching and chimney method and coil embolization procedure

Ann Vasc Dis. 2013;6(3):601-11. doi: 10.3400/avd.cr.13-00070. Epub 2013 Sep 5.

Abstract

Objective: We here describe our experience with innovative uses of these devices.

Patients and methods: We reviewed treatment outcomes of 310 endovascular abdominal aortic repair (EVAR) and 83 thoracic endovascular aortic repair (TEVAR) cases performed between August 2007 and February 2012. We separately assessed results in elderly and high-risk patients who had a novel procedure. This group included 94 patients who underwent EVAR with IIA embolization, 10 patients who had EVAR and a renal artery chimney procedure for a short aortic neck, 20 patients who had two de-branching TEVAR or Chimney method for thoracic aortic aneurysms (TAA) and 3 patients who had debranching TEVAR for thoracic abdominal aortic aneurysms (TAAA).

Results: Of the 393 patients given stent grafts (SGs), 3 (0.8%) died in the hospital, including 1 patient with pneumonia who underwent EVAR and IIA embolization and 1 patient with a cerebral infarction who had TEVAR. Four patients (4.3%) who were treated with EVAR with internal iliac artery (IIA) embolization presented with residual buttock claudication 6 months postoperatively, and 3 patients (3.2%) had onset of ischemic enteritis; however, in all 7 patients, the condition resolved without additional intervention. In the 10 patients who had EVAR and a renal artery chimney method, the landing zone (LZ) was ≤10 mm, but neither endoleak nor renal artery occlusion was observed perioperatively or during midterm follow-up. Of the 20 patients who had a 2-debranching TEVAR, including 9 in whom the chimney method was used with the LZ in zone 0, 1 (5%) had a residual endoleak. In 3 patients with TAAA, we used SGs to cover 4 abdominal branches and bypassed the visceral artery; the outcomes were good, with all patients being ambulatory at hospital discharge.

Conclusion: Among innovative SGs treatments, the debranching procedure and the chimney method using catheterization and the coil-embolization technique provided good outcomes, as used in addition to surgical procedures. Aortic aneurysm treatment will become increasingly noninvasive with the continuing development of more innovative ways to use the SGs currently available in Japan. (English Translation of Jpn J Vasc Surg 2012; 21: 165-173).

Keywords: aortic aneurysm; chimney method; coil embolization; debranching procedure; stentgraft.