Towards real-time in situ polyp detection in WCE images using a boosting-based approach

Annu Int Conf IEEE Eng Med Biol Soc. 2013:2013:5711-4. doi: 10.1109/EMBC.2013.6610847.

Abstract

This paper presents a new embeddable method for polyp detections in Wireless Capsule Endoscopic - WCE images. this approach consists first of extracting candidate polyps within the image using geometric considerations about related shape, and second, in classifying (polyp/non-polyp) obtained candidates by a boosting-based method using texture features. The proposed approach has been designed in accordance with the hardware constraints related to FPGA implementation for integration within WCE imaging device. The classification performance of the method have been evaluated on a large dataset of 300 polyps, and 1200 non-polyps images. Experiments show interesting and promising performance: the boosting-based classification is characterized by a sensitivity of 91%, a specificity of 95% and a false detection rate of 4.8%, the detection rate of the overall processing chain being of 68%. The performance of the boosting-based classification are in accordance with the most recent reference on this particular topic using the same dataset. Building of a dedicated WCE image database should permit the improvement of the global detection rate.

MeSH terms

  • Algorithms*
  • Capsule Endoscopy / methods*
  • Computer Systems*
  • Databases as Topic
  • Humans
  • Image Processing, Computer-Assisted
  • Polyps / diagnosis*
  • Sensitivity and Specificity