Proposal on 3-D endoscope by using grid-based active stereo

Annu Int Conf IEEE Eng Med Biol Soc. 2013:2013:5694-7. doi: 10.1109/EMBC.2013.6610843.

Abstract

In this paper, we propose a novel 3-D endoscope system by using grid-based active stereo. In the proposed system, projection of a waved-grid pattern that consists of vertical and horizontal sinusoidal lines realizes accurate shape acquisition in sub-pixel accuracy. We develop a small pattern projector implementable to a head of a ready-made endoscope, and examine 3-D shape reconstruction by actual equipment. As the result of the measurement for a known-shaped object, which is a hexagonal cylinder, the error of length measurement is below 0.9% and the error of angle measurement is below 2.2%. We make a measurement of animal organ meat, and confirm that the system can reconstruct a 3-D shape of the organ surface.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Animals
  • Calibration
  • Endoscopes*
  • Endoscopy / instrumentation*
  • Imaging, Three-Dimensional / instrumentation
  • Imaging, Three-Dimensional / methods*
  • Meat
  • Organ Size
  • Pattern Recognition, Automated