An optically coupled system for quantitative monitoring of MRI gradient currents induced into endocardial leads

Annu Int Conf IEEE Eng Med Biol Soc. 2013:2013:2400-3. doi: 10.1109/EMBC.2013.6610022.

Abstract

The time-varying gradient fields generated during Magnetic Resonance Imaging (MRI) procedures have the potential to induce electrical current on implanted endocardial leads. Whether this current can result in undesired cardiac stimulation is unknown. This paper presents an optically coupled system with the potential to quantitatively measure the currents induced by the gradient fields into endocardial leads during MRI procedures. Our system is based on a microcontroller that works as analog-to-digital (A/D) converter and sends the current signal acquired from the lead to an optical high-speed light-emitting-diode transmitter. Plastic fiber guides the light outside the MRI chamber, to a photodiode receiver and then to an acquisition board connected to a PC. The preliminary characterization of the performances of the system is also presented.

MeSH terms

  • Artifacts
  • Defibrillators, Implantable*
  • Humans
  • Magnetic Fields
  • Magnetic Resonance Imaging
  • Optical Devices
  • Pacemaker, Artificial*