Safe magnetic resonance imaging scanning of patients with cardiac rhythm devices: a role for computer modeling

Heart Rhythm. 2013 Dec;10(12):1815-21. doi: 10.1016/j.hrthm.2013.10.009. Epub 2013 Oct 4.

Abstract

Background: Although there are several hazards for patients with implanted pacemakers and defibrillators in the magnetic resonance imaging (MRI) environment, evaluation of lead electrode heating is the most complex because of the many influencing variables: patient size, anatomy, body composition, patient position in the bore, scan sequence (radiofrequency power level), lead routing, and lead design. Although clinical studies are an important step in demonstrating efficacy, demonstrating safety through clinical trials alone is not practical because of this complexity.

Objective: The purpose of this study was to develop a comprehensive modeling framework to predict the probability of pacing capture threshold (PCT) change due to lead electrode heating in the MRI environment and thus provide a robust safety evaluation.

Methods: The lead heating risk was assessed via PCT change because this parameter is the most clinically relevant measure of lead heating. The probability for PCT change was obtained by combining the prediction for power at the electrode-tissue interface obtained via simulations with a prediction for PCT change as a function of radiofrequency power obtained via an in vivo canine study.

Results: The human modeling framework predicted that the probability of a 0.5-V PCT change due to an MRI scan for the Medtronic CapSureFix MRI SureScan model 5086 MRI leads is <1/70,000 for chest scans and <1/10,000,000 for either head scans or lower torso scans.

Conclusion: The framework efficiently models millions of combinations, delivering a robust evaluation of the lead electrode heating hazard. This modeling approach provides a comprehensive safety evaluation that is impossible to achieve using phantom testing, animal studies, or clinical trials alone.

Keywords: 5086 MRI lead; AAMI; Association for the Advancement of Medical Instrumentation; Computer modeling; Hazards; ISO; International Organization for Standardization; Lead electrode heating; MR conditional; MRI; Magnetic resonance imaging; PCT; RF; SureScan; magnetic resonance imaging; pacing capture threshold; radiofrequency.

Publication types

  • Comparative Study

MeSH terms

  • Animals
  • Arrhythmias, Cardiac / diagnosis
  • Arrhythmias, Cardiac / therapy*
  • Computer Simulation*
  • Disease Models, Animal
  • Dogs
  • Equipment Design
  • Equipment Safety
  • Humans
  • Imaging, Three-Dimensional
  • Magnetic Resonance Imaging, Cine / methods*
  • Pacemaker, Artificial*
  • Phantoms, Imaging
  • Reproducibility of Results