ECAP spread of excitation with virtual channels and physical electrodes

Hear Res. 2013 Dec:306:93-103. doi: 10.1016/j.heares.2013.09.014. Epub 2013 Oct 3.

Abstract

The primary goal of this study was to evaluate physiological spatial excitation patterns for stimulation of adjacent physical electrodes and intermediate virtual channels. Two experiments were conducted that utilized electrically evoked compound action potential (ECAP) spread-of-excitation (SOE) functions obtained with the traditional forward-masking subtraction method. These two experiments examined spatial excitation patterns for virtual-channel maskers and probes, respectively. In Experiment 1, ECAP SOE patterns were obtained for maskers applied to physical electrodes and virtual channels to determine whether virtual-channel maskers yield SOE patterns similar to those predicted from physical electrodes. In Experiment 2, spatial separation of SOE functions was compared for two adjacent physical probe electrodes and the intermediate virtual channel to determine the extent to which ECAP SOE patterns for virtual-channel probes are spatially separate from those obtained with physical electrodes. Data were obtained for three electrode regions (basal, middle, apical) for 35 ears implanted with Cochlear (N = 16) or Advanced Bionics (N = 19) devices. Results from Experiment 1 showed no significant difference between predicted and measured ECAP amplitudes for Advanced Bionics subjects. Measured ECAP amplitudes for virtual-channel maskers were significantly larger than the predicted amplitudes for Cochlear subjects; however, the difference was <2 μV and thus is likely not clinically significant. Results from Experiment 2 showed that the probe set in the apical region demonstrated the least amount of spatial separation amongst SOE functions, which may be attributed to more uniform nerve survival patterns, closer electrode spacing, and/or the tapered geometry of the cochlea. As expected, adjacent physical probes demonstrated greater spatial separation than for comparisons between each physical probe and the intermediate virtual channel. Finally, the virtual-channel SOE functions were generally weighted toward the basal electrode in the pair.

Keywords: AB; Advanced Bionics; BEDCS; Bionic Ear Data Collection System; CI; CL; CPI II; Clinical Programming Interface; ECAP; MPI; NRT; Neural Response Telemetry; PM; PSP; Platinum Series Processor; RM ANOVA; SD; SOE; VM; cochlear implant; current level units; electrically evoked compound action potential; masker-probe interval; physical-electrode maskers; pps; pulses per second; repeated-measures analysis of variance; spread of excitation; standard deviation; virtual-channel maskers.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Action Potentials / physiology*
  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Bionics*
  • Child
  • Child, Preschool
  • Cochlear Implantation / methods*
  • Cochlear Implants*
  • Electric Impedance
  • Electric Stimulation
  • Electrodes*
  • Equipment Design
  • Humans
  • Infant
  • Middle Aged
  • Neurons / pathology
  • Signal Processing, Computer-Assisted
  • Telemetry
  • Young Adult