Field evaluation in Chad of community usage of CD4 T lymphocyte counting by alternative single-platform flow cytometry

BMC Health Serv Res. 2013 Oct 1:13:373. doi: 10.1186/1472-6963-13-373.

Abstract

Background: Field and community evaluation of the routine usage of CD4 T counting platforms is essential in resource-poor countries for efficient and cost-effective monitoring of HIV-infected adults and children attending health care centers.

Methods: We herein addressed the principal issues raised by the implementation of the single-platform, volumetric Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK) in 8 community HIV monitoring laboratories of different levels throughout Chad. This is a country with particularly difficult conditions, both in terms of climate and vast geographical territory, making the decentralization of the therapeutic management of HIV-infected patients challenging.

Results: The routine usage of the Auto40 flow cytometers for a period of 5 years (2008-2013) confirms the reliability and robustness of the analyzer for community-based CD4 T cell enumeration in terms of both absolute numbers and percentages to enable accurate monitoring of HIV-infected adults and children. However, our observations suggest that the Auto40 mini flow cytometer is not suitable for all laboratories as it is oversized and ultimately very expensive.

Conclusion: The Chad experience with the Auto40 flow cytometer suggests that its usage in resource-limited settings should be mainly reserved to reference (level 1) or district (level 2) laboratories, rather than to laboratories of health care centres (level 3).

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • CD4 Lymphocyte Count / instrumentation
  • CD4 Lymphocyte Count / methods*
  • Chad
  • Child
  • Flow Cytometry / instrumentation
  • Flow Cytometry / methods*
  • HIV Infections / diagnosis
  • Humans