Detection of C-reactive protein based on magnetic nanoparticles and capillary zone electrophoresis with laser-induced fluorescence detection

J Chromatogr A. 2013 Nov 8:1315:188-94. doi: 10.1016/j.chroma.2013.09.042. Epub 2013 Sep 17.

Abstract

A simple and fast method based on magnetic nanoparticles (MNPs) and capillary zone electrophoresis (CZE) with laser-induced fluorescence (LIF) detection was developed for the detection of C-reactive protein (CRP). To optimize the CZE conditions, several factors including buffer compositions, buffer ionic strength, buffer pH, applied voltage and capillary temperature have been examined. The optimal separation buffer selected was a 30 mM sodium phosphate (PB) buffer, pH 8.0. The optimal CE applied voltage and temperature selected were 20 kV and 35°C, respectively. The CZE profile of the MNP-1°Ab-CRP-2°Ab/FITC bioconjugates showed good reproducibility. One major peak was observed for the MNP bioconjugates. The quantitative analysis also showed good results. The coefficient of variation (CV%) for the major peak area was 8.7%, and the CV% for the major peak migration time was 2.5%. The linear range for CRP analysis was 10-150 μg/mL, and the concentration limit of detection (LOD) was 9.2 μg/mL. Non-specific interactions between bovine serum albumin (BSA) and the system can be prevented by including 10% (v/v) of human plasma in the binding buffers. The CE/LIF method might be helpful for analyzing high concentrations of CRP in a patient's plasma after an acute-phase inflammation. This new method demonstrated the possibility of using MNPs and CE/LIF for the detection of proteins, and provided information for the establishment of appropriate CE conditions.

Keywords: C-reactive protein; Capillary electrophoresis; Fluorescence; Magnetic nanoparticle.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • C-Reactive Protein / analysis*
  • C-Reactive Protein / chemistry
  • Electrophoresis, Capillary / methods*
  • Humans
  • Hydrogen-Ion Concentration
  • Limit of Detection
  • Linear Models
  • Magnetite Nanoparticles / chemistry*
  • Osmolar Concentration
  • Particle Size
  • Reproducibility of Results
  • Spectrometry, Fluorescence / methods*

Substances

  • Magnetite Nanoparticles
  • C-Reactive Protein