Hybrid-SPRITE MRI

J Magn Reson. 2013 Oct:235:6-14. doi: 10.1016/j.jmr.2013.07.003. Epub 2013 Jul 18.

Abstract

In a FID based frequency encoding MRI experiment the central part of k-space is not generally accessible due to the probe dead time. This portion of k-space is however crucial for image reconstruction. SPRITE (Single Point Ramped Imaging with T1 Enhancement), SPI with a linearly ramped phase encode gradient, has been employed to image short relaxation time systems for many years with great success. It is a robust imaging method in significant measure because it provides acquisition of high quality k-space origin data. We propose a new sampling scheme, termed hybrid-SPRITE, combining phase and frequency encoding to ensure high quality images with reduced acquisition times, reduced gradient duty cycle and increased sensitivity. In hybrid-SPRITE, numerous time domain points are collected to assist image reconstruction. An Inverse Non-uniform Discrete Fourier Transform (INDFT) is employed in 1D applications. A pseudo-polar grid is exploited in 2D hybrid-SPRITE for rapid and accurate image reconstruction.

Keywords: Frequency and phase encoding; Inverse Non-uniform Discrete Fourier Transform; Profiles; Pseudo-polar grid; SPRITE; Sensitivity.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Fourier Analysis
  • Humans
  • Image Enhancement / methods
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Spectroscopy
  • Phantoms, Imaging
  • Radio Waves