Online in vivo dosimetry in high dose rate prostate brchytherapy with MOSkin detectors: in phantom feasibility study

Appl Radiat Isot. 2014 Jan:83 Pt C:222-6. doi: 10.1016/j.apradiso.2013.06.001. Epub 2013 Jun 12.

Abstract

MOSkin detectors were studied to perform real-time in vivo dose measurements in high dose rate prostate brachytherapy. Measurements were performed inside an urethral catheter in a gel phantom simulating a real prostate implant. Measured and expected doses were compared and the discrepancy was found to be within 8.9% and 3.8% for single MOSkin and dual-MOSkin configurations, respectively. Results show that dual-MOSkin detectors can be profitably adopted in prostate brachytherapy treatments to perform real-time in vivo dosimetry inside the urethra.

Keywords: High dose rate; In vivo dosimetry; MOSFET; Prostate brachytherapy.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brachytherapy*
  • Feasibility Studies
  • Humans
  • Male
  • Phantoms, Imaging*
  • Prostatic Neoplasms / radiotherapy*
  • Radiotherapy Dosage*