An electroanalytical approach for evaluation of biochar adsorption characteristics and its application for lead and cadmium determination

Bioresour Technol. 2013 Sep:143:40-5. doi: 10.1016/j.biortech.2013.05.107. Epub 2013 May 31.

Abstract

This work describes for first time the use of electroanalytical techniques for evaluation of adsorptive proprieties of biochar using it as electrode modifier and its application for preconcentration and determination of Lead(II) and Cadmium(II) under differential pulse adsorptive voltammetric conditions (DPAdSV). Samples of biochars were obtained from castor oil cake using a predefined set of experimental conditions varying the heating rate (V), final temperature (T) and warm-up period (P) and subsequently used for carbon paste modified electrode (CPME) preparation. The proposed method was applied for Lead(II) and Cadmium(II) determination in spiked simulated industrial effluents and the limit of detection obtained for both metals were adequated for determination of these evaluated ions taking into account the limits established by Brazilian legislation. For all samples analyzed, recoveries ranged from 95% to 104% were obtained and no significative interferences were observed for common cations in water samples.

Keywords: Adsorption of Lead and Cadmium; Biochar; Carbon paste electrode; Stripping analysis.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adsorption
  • Cadmium / analysis*
  • Charcoal*
  • Electrochemical Techniques / methods*
  • Lead / analysis*
  • Limit of Detection

Substances

  • biochar
  • Cadmium
  • Charcoal
  • Lead