Auto-aligning stimulated emission depletion microscope using adaptive optics

Opt Lett. 2013 Jun 1;38(11):1860-2. doi: 10.1364/OL.38.001860.

Abstract

Stimulated emission depletion (STED) microscopy provides diffraction-unlimited resolution in fluorescence microscopy. Imaging at the nanoscale, however, requires precise alignment of the depletion and excitation laser foci of the STED microscope. We demonstrate here that adaptive optics can be implemented to automatically align STED and confocal images with a precision of 4.3 ± 2.3 nm.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Image Processing, Computer-Assisted
  • Lasers
  • Microscopy, Fluorescence / methods*
  • Optical Phenomena*