Long-time quality assessment of the Elecsys Troponin T hs assay

Clin Biochem. 2013 Aug;46(12):1055-1057. doi: 10.1016/j.clinbiochem.2013.03.022. Epub 2013 Apr 9.

Abstract

Recently, Roche announced a problem with the calibrators for their high sensitive cardiac troponin T (hscTnT) assay. We have monitored the performance of the assay since its introduction into clinical use in December 2009 and we're in a unique position to provide quantitative information concerning the effects of this problem. Our data document that the measured cTnT concentration in the hscTnT assay dropped by 5.8 ng/L around the 99th percentile before its recalibration in May 2012. Thus, the hscTnT levels in our hospital have been underestimated around this cut-off value since its introduction. The approach we used may be one that other laboratories should consider.

Keywords: Calibration; High sensitive troponin assays.

MeSH terms

  • Calibration
  • Chemistry, Clinical / methods*
  • Humans
  • Quality Assurance, Health Care*
  • Sensitivity and Specificity
  • Time Factors
  • Troponin T / blood*