Analysis of nucleosides and nucleotides in infant formula by liquid chromatography-tandem mass spectrometry

Anal Bioanal Chem. 2013 Jun;405(15):5311-9. doi: 10.1007/s00216-013-6935-9. Epub 2013 Apr 5.

Abstract

A method for the simultaneous analysis of nucleosides and nucleotides in infant formula using reversed-phase liquid chromatography-tandem mass spectrometry is described. This approach is advantageous for compliance testing of infant formula over other LC-MS methods in which only nucleotides or nucleosides are measured. Following sample dissolution, protein was removed by centrifugal ultrafiltration. Chromatographic analyses were performed using a C18 stationary phase and gradient elution of an ammonium acetate/bicarbonate buffer, mass spectrometric detection and quantitation by a stable isotope-labelled internal standard technique. A single laboratory validation was performed, with spike recoveries of 80.1-112.9% and repeatability relative standard deviations of 1.9-7.2%. Accuracy as bias was demonstrated against reference values for NIST1849a certified reference material. The method has been validated for the analysis of bovine milk-based, soy-based, caprine milk-based and hydrolysed milk protein-based infant formulae.