Comparison of three different commercial PCR assays for the detection of pathogens in critically ill sepsis patients

Med Klin Intensivmed Notfmed. 2013 May;108(4):311-8. doi: 10.1007/s00063-013-0227-1. Epub 2013 Mar 22.

Abstract

Introduction: The high mortality rate associated with sepsis necessitates a timely identification of the causative organism in order to optimize antimicrobial therapy. PCR assays are increasingly being used for this purpose. The aim of this study was to compare three commercially available PCR systems for the diagnosis of systemic infections.

Patients and methods: In a prospective observational study, a broad-range (SepsiTest®; Molzym, Bremen, Germany) and two multiplex PCR assays (VYOO®; SIRS-Lab, Jena, Germany and LightCycler® SeptiFast; Roche, Mannheim, Germany) were compared to blood cultures with respect to the clinical course of 50 critically ill patients with sepsis, severe sepsis or septic shock.

Results: Pathogens were detected by PCR in 12 % (SepsiTest®), 10 % (VYOO®) and 14 % (LightCycler® SeptiFast) of samples and in 26 % by blood culture. Negative results were obtained using all four methods in 32 samples (64 %) and 3 (6 %) samples were positive in all tests. Upon consideration of additional diagnostic findings and the clinical course, eight (16 %) of the positive blood culture results were deemed clinically relevant. All three PCR assays could also identify the causative organism (or a specific gene thereof) in three of these eight positive blood cultures, whereas for five of the eight, all three PCR assays were negative. In one patient with a negative blood culture, the SepsiTest®, VYOO® and LightCycler® SeptiFast assays were positive for Streptococcus species. The PCR assays appeared to be less susceptible than blood cultures to false-positive results arising from contamination with coagulase-negative staphylococcal organisms.

Conclusion: There was some variability between the three PCR assays tested and the corresponding blood cultures with regards to the type of pathogen detected. The three PCR assays appeared to be less susceptible to false-positive results than blood cultures.

Publication types

  • Comparative Study
  • Observational Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Anti-Bacterial Agents / therapeutic use
  • Bacterial Infections / diagnosis*
  • Bacterial Infections / drug therapy
  • Bacterial Infections / microbiology*
  • Bacteriological Techniques
  • Blood / microbiology
  • Critical Care*
  • Female
  • Germany
  • Humans
  • Male
  • Middle Aged
  • Multiplex Polymerase Chain Reaction / methods*
  • Polymerase Chain Reaction / methods*
  • Predictive Value of Tests
  • Prospective Studies
  • Sepsis / diagnosis*
  • Sepsis / drug therapy
  • Sepsis / microbiology*

Substances

  • Anti-Bacterial Agents