Invasive hemodynamic characteristics of low gradient severe aortic stenosis despite preserved ejection fraction

J Am Coll Cardiol. 2013 Apr 30;61(17):1799-808. doi: 10.1016/j.jacc.2013.02.009. Epub 2013 Feb 28.

Abstract

Objectives: The study sought to compare echocardiographic with invasive hemodynamic data in patients with "paradoxic" aortic stenosis and in patients with conventionally defined severe aortic stenosis.

Background: Controversy exists whether low gradient severe aortic stenosis despite preserved ejection fraction ("paradoxic" aortic stenosis; aortic valve area <1 cm(2), mean gradient <40 mm Hg, ejection fraction >50%), which has been mainly diagnosed by echocardiography (echo), may be largely due to mistakes in echocardiographic measurements.

Methods: We compared echocardiographic and invasive hemodynamic data from 58 patients (43% male, mean age 77 ± 5 years) with "paradoxic" aortic stenosis. Data of 22 patients (45% male, mean age 73 ± 7 years) with conventionally defined severe aortic stenosis area (aortic valve area ≤1 cm(2), mean gradient >40 mm Hg, ejection fraction ≥50%) were also analyzed.

Results: In patients with "paradoxic" aortic stenosis, orifice area by echo (0.80 ± 0.15 cm(2)) and catheterization showed modest agreement, whether stroke volume was measured by oxymetry (0.69 ± 0.16 cm(2), bias 0.14 ± 0.17 cm(2)), or by thermodilution (0.85 ± 0.19 cm(2), bias -0.03 ± 0.19 cm(2)). Mean systolic gradients were very similar (32 ± 7 mm Hg vs. 31 ± 6 mm Hg; bias -0.08 ± 7.8 mm Hg). In comparison, in patients with conventionally defined severe aortic stenosis, orifice area by echo was 0.72 ± 0.17 cm(2) and by catheterization 0.51 ± 0.15 cm(2) (oxymetry) and 0.68 ± 0.21 cm(2) (thermodilution), respectively, and mean systolic gradient 51 ± 10 mm Hg and 55 ± 8 mm Hg, respectively. Ejection fractions did not differ significantly in both groups. Ascending aortic diameter was significantly smaller in the "paradoxic" aortic stenosis group than in patients with conventionally defined severe aortic stenosis (28 ± 5 mm vs. 31 ± 5 mm), and energy loss index was significantly larger (0.51 ± 0.12 cm(2)/m(2) vs. 0.42 ± 0.09 cm(2)/m(2), respectively). Heart rate and mean blood pressure during echo and catheterization were not significantly different.

Conclusions: Occurrence of low gradient severe aortic stenosis despite preserved ejection fraction was confirmed by invasive hemodynamics and was not the result of a systematic bias in the echo calculation of aortic orifice area.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve Stenosis / diagnostic imaging*
  • Aortic Valve Stenosis / physiopathology*
  • Aortic Valve Stenosis / surgery
  • Cardiac Catheterization*
  • Echocardiography
  • Female
  • Hemodynamics
  • Humans
  • Male
  • Middle Aged
  • Severity of Illness Index
  • Stroke Volume*
  • Vascular Resistance