The effects of early osseointegration in different implant sites in rabbit tibias

J Mater Sci Mater Med. 2013 Apr;24(4):959-65. doi: 10.1007/s10856-012-4847-4. Epub 2013 Feb 21.

Abstract

The aim of this study was to evaluate the early osseointegration of implants with the same surface treatment in different implant sites in rabbit tibias after 4 weeks. A total of 42 acid-etched implants were implanted in three different sites in the tibia: group A was 2.08 ± 0.18 mm below epiphyseal line; group B was 7.00 ± 0.61 mm below the epiphyseal line; group C was 13.01 ± 1.26 mm below the epiphyseal line. After 4 weeks, the average bone-to-implant contact (BIC) values were as follows: group A, 40.02 ± 4.82 %; group B, 28.20 ± 5.41 %; group C, 20.76 ± 3.10 %. The BIC measurements yielded statistically significant differences among group A, group B and group C (P < 0.01); group A demonstrated the best osseointegration. In the present study, the different implantation sites in the selected 20-mm area demonstrated different early osseointegration; the sites located 7 ± 1.5 mm below the epiphyseal line were best suited for observing the effectiveness of early osseointegration among the three sites. The statistical results of the early osseointegration of implants are therefore affected by the location of the implant sites in this 20-mm area.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Male
  • Osseointegration*
  • Prostheses and Implants*
  • Rabbits
  • Surface Properties
  • Tibia*