Heat bump on a monochromator crystal measured with X-ray grating interferometry

J Synchrotron Radiat. 2013 Mar;20(Pt 2):300-5. doi: 10.1107/S0909049513001817. Epub 2013 Feb 7.

Abstract

Deformation of the first crystal of an X-ray monochromator under the heat load of a high-power beam, commonly referred to as `heat bump', is a challenge frequently faced at synchrotron beamlines. Here, quantitative measurements of the deformations of an externally water-cooled silicon (111) double-crystal monochromator tuned to a photon energy of 17.6 keV are reported. These measurements were made using two-dimensional hard X-ray grating interferometry, a technique that enables in situ at-wavelength wavefront investigations with high angular sensitivity. The observed crystal deformations were of the order of 100 nm in the meridional and 5 nm in the sagittal direction, which lead to wavefront slope errors of up to 4 µrad in the meridional and a few hundred nanoradians in the sagittal direction.

Keywords: double-crystal monochromator; grating interferometry; heat bump; wavefront characterization.