Static and cyclic performance evaluation of sensors for human interface pressure measurement

Annu Int Conf IEEE Eng Med Biol Soc. 2012:2012:162-5. doi: 10.1109/EMBC.2012.6345896.

Abstract

Researchers and clinicians often desire to monitor pressure distributions on soft tissues at interfaces to mechanical devices such as prosthetics, orthotics or shoes. The most common type of sensor used for this type of applications is a Force Sensitive Resistor (FSR) as these are convenient to use and inexpensive. Several other types of sensors exist that may have superior sensing performance but are less ubiquitous or more expensive, such as optical or capacitive sensors. We tested five sensors (two FSRs, one optical, one capacitive and one fluid pressure) in a static drift and cyclic loading configuration. The results show that relative to the important performance characteristics for soft tissue pressure monitoring (i.e. hysteresis, drift), many of the sensors tested have significant limitations. The FSRs exhibited hysteresis, drift and loss of sensitivity under cyclic loading. The capacitive sensor had substantial drift. The optical sensor had some hysteresis and temperature-related drift. The fluid pressure sensor performed well in these tests but is not as flat as the other sensors and is not commercially available. Researchers and clinicians should carefully consider the convenience and performance trade-offs when choosing a sensor for soft-tissue pressure monitoring.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Electric Capacitance
  • Humans
  • Models, Theoretical*
  • Monitoring, Physiologic / instrumentation*
  • Monitoring, Physiologic / methods*
  • Optical Devices
  • Pressure*