Sensitive HPLC-PDA determination of tamoxifen and its metabolites N-desmethyltamoxifen, 4-hydroxytamoxifen and endoxifen in human plasma

J Pharm Biomed Anal. 2013 Mar 25:76:13-20. doi: 10.1016/j.jpba.2012.12.005. Epub 2012 Dec 13.

Abstract

A highly sensitive HPLC-UV method for the simultaneous determination of tamoxifen, N-desmethyltamoxifen, 4-hydroxytamoxifen and endoxifen in human plasma samples was developed and validated. The method employs a two step liquid-liquid extraction and a reversed phase separation on a Hypersil Gold(®) C18 column (150mm×4.6mm, 5μm) with isocratic elution. Mobile phase was a mixture of triethylammonium phosphate buffer 5mM pH 3.3 and acetonitrile (57:43, v/v). Total analytical run time was 16min. Precision assays showed CV % lower than 10.53% and accuracy in the range of 93.0-104.2%. The lower limits of quantification (0.75-8.5ngml(-1)) are adequate to measure clinically relevant concentrations in plasma samples. The method was successfully applied to 110 clinical plasma samples. Median plasma levels and interquartile range were: tamoxifen 55.77ngml(-1) (38.42-83.69ngml(-1)), N-desmethyltamoxifen 124.83ngml(-1) (86.81-204.80ngml(-1)), 4-hydroxytamoxifen 1.09ngml(-1) (0.76-1.53ngml(-1)) and endoxifen 6.18ngml(-1) (4.17-8.22ngml(-1)). The procedure has adequate analytical performance and can be employed in therapeutic drug monitoring of tamoxifen or pharmacokinetics studies.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Antineoplastic Agents, Hormonal / blood*
  • Antineoplastic Agents, Hormonal / metabolism
  • Chromatography, High Pressure Liquid / methods*
  • Drug Monitoring / methods
  • Female
  • Humans
  • Limit of Detection
  • Sensitivity and Specificity
  • Tamoxifen / analogs & derivatives
  • Tamoxifen / blood*
  • Tamoxifen / metabolism

Substances

  • Antineoplastic Agents, Hormonal
  • Tamoxifen
  • afimoxifene
  • 4-hydroxy-N-desmethyltamoxifen
  • N-desmethyltamoxifen