Magnetic particle detection (MPD) for in-vitro dosimetry

Biosens Bioelectron. 2013 May 15:43:88-93. doi: 10.1016/j.bios.2012.12.011. Epub 2012 Dec 8.

Abstract

In-vitro tests intended for evaluating the potential health effects of magnetic nanoparticles generally require an accurate measure of cell dose to promote the consistent use and interpretation of biological response. Here, a simple low-cost inductive sensor is developed for quickly determining the total mass of magnetic nanoparticles that is bound to the plasma membrane and internalized by cultured cells. Sensor operation exploits an oscillating magnetic field (f0=250kHz) together with the nonlinear response of particle magnetization to generate a harmonic signal (f3=750kHz) that varies linearly with particulate mass (R(2)>0.999) and is sufficiently sensitive for detecting ∼100ng of carboxyl-coated iron-oxide nanoparticles in under a second. When exploited for measuring receptor-mediated nanoparticle uptake in RAW 264.7 macrophages, results show that the achieved dosimetric performance is comparable with relatively expensive analytical techniques that are much more time-consuming and labor-intensive to perform. The described sensing is therefore potentially better suited for low-cost in-vitro assays that require fast and quantitative magnetic particle detection.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Biological Assay / instrumentation*
  • Biosensing Techniques / instrumentation*
  • Cell Membrane / chemistry
  • Cell Membrane / radiation effects*
  • Equipment Design
  • Equipment Failure Analysis
  • Magnetics / instrumentation*
  • Magnetite Nanoparticles / analysis*
  • Magnetite Nanoparticles / chemistry
  • Magnetite Nanoparticles / radiation effects*
  • Mice
  • Radiometry / instrumentation*

Substances

  • Magnetite Nanoparticles