A first-in-man study of sirolimus-eluting, biodegradable polymer coated cobalt chromium stent in real life patients

Indian Heart J. 2012 Nov-Dec;64(6):547-52. doi: 10.1016/j.ihj.2012.07.011. Epub 2012 Jul 27.

Abstract

Introduction: Despite considerable benefits associated with current drug-eluting stents, continued attention to the safety, efficacy, and deliverability of available drug-eluting stent has led to the development of newer stent.

Methods: This study was a single-centre, prospective, non-randomized, first-in-man study which included clinical follow-up data was collected at 1, 8 and 12 months after the procedure. The study included 105 patients with de novo native coronary artery lesions including multi-vessel disease treated with Supralimus-Core® stent. Repeat angiography was performed 8 months post-stent implantation.

Results: At quantitative coronary angiography 8-month luminal late loss was 0.39 ± 0.33 mm in-stent and 0.33 ± 0.35 mm in-segment. The incidence of any major adverse cardiac event at 30 days, 8 months and 12 months was 1 (1%), 6 (6%) and 7 (7%) respectively.

Conclusion: This study demonstrates that the Supralimus-Core® SES is a safe and effective treatment for patients with obstructive coronary artery disease. ClinicalTrials.gov ID: NCT00811616.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorbable Implants
  • Chromium
  • Coated Materials, Biocompatible
  • Cobalt
  • Coronary Angiography
  • Coronary Disease / diagnostic imaging
  • Coronary Disease / therapy*
  • Drug-Eluting Stents*
  • Female
  • Humans
  • Immunosuppressive Agents / administration & dosage*
  • Male
  • Middle Aged
  • Polymers
  • Prospective Studies
  • Sirolimus / administration & dosage*
  • Treatment Outcome

Substances

  • Coated Materials, Biocompatible
  • Immunosuppressive Agents
  • Polymers
  • Chromium
  • Cobalt
  • Sirolimus

Associated data

  • ClinicalTrials.gov/NCT00811616