High field carotid vessel wall imaging: a study on reproducibility

Eur J Radiol. 2013 Apr;82(4):680-5. doi: 10.1016/j.ejrad.2012.11.028. Epub 2012 Dec 14.

Abstract

Purpose: Currently, a multi-contrast protocol, including a combination of five MR-sequences is used as reference standard for morphologic imaging and quantitative measurements of the carotid artery vessel wall. The purpose of this study is to investigate the scan-rescan reproducibility together with intra- and inter-observer reproducibility of each of the five MR-sequences.

Methods: Twenty healthy volunteers (55% male, mean age=26 years) underwent repeated MR-examinations (3T-Philips-MRI) of the left carotid artery vessel wall with five sequences; T1-TFE, T2-TSE, PD-TSE, T1-TSE and 3D TOF. A standard phased-array coil with two flexible elements of 14cm×17cm was used to obtain nine transverse imaging sections of the left carotid artery with identical in-plane resolution (0.46mm×0.46mm). Reproducibility analysis was performed in 3 slices of the common carotid artery for all sequences.

Results: For, scan-rescan reproducibility, intra class correlation coefficients (ICC) were excellent for all sequences and ranged from 0.79 to 0.95. The intra-observer ICC ranged from 0.89 to 0.98 and the inter-observer ICC ranged from 0.84 to 0.96, for both lumen and vessel wall assessment.

Conclusions: By high field MR imaging, vessel wall and lumen area of the carotid artery can be assessed with excellent scan-rescan, intra- and inter-observer reproducibility for all five sequences.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Carotid Arteries / anatomy & histology*
  • Female
  • Humans
  • Image Interpretation, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Male
  • Reproducibility of Results