Calculation of physiologic dead space: comparison of ventilator volumetric capnography to measurements by metabolic analyzer and volumetric CO2 monitor

Respir Care. 2013 Jul;58(7):1143-51. doi: 10.4187/respcare.02116. Epub 2012 Dec 4.

Abstract

Background: Calculation of physiologic dead space (dead space divided by tidal volume [VD/VT]) using the Enghoff modification of the Bohr equation requires measurement of the partial pressure of mean expired CO2 (PĒCO2) by exhaled gas collection and analysis, use of a metabolic analyzer, or use of a volumetric CO2 monitor. The Dräger XL ventilator is equipped with integrated volumetric CO2 monitoring and calculates minute CO2 production (VCO2). We calculated PĒCO2 and VD/VT from ventilator derived volumetric CO2 measurements of VCO2 and compared them to metabolic analyzer and volumetric CO2 monitor measurements.

Methods: A total of 67 measurements in 36 subjects recovering from acute lung injury or ARDS were compared. Thirty-one ventilator derived measurements were compared to measurements using 3 different metabolic analyzers, and 36 ventilator derived measurements were compared to measurements from a volumetric CO2 monitor.

Results: There was a strong agreement between ventilator derived measurements and metabolic analyzer or volumetric CO2 monitor measurements of PĒCO2 and VD/VT. The correlations, bias, and precision between the ventilator and metabolic analyzer measurements for PĒCO2 were r = 0.97, r(2) = 0.93 (P < .001), bias -1.04 mm Hg, and precision ± 1.47 mm Hg. For VD/VT the correlations were r = 0.95 and r(2) = 0.91 (P < .001), and the bias and precision were 0.02 ± 0.04. The correlations between the ventilator and the volumetric CO2 monitor for PĒCO2 were r = 0.96 and r(2) = 0.92 (P < .001), and the bias and precision were -0.19 ± 1.58 mm Hg. The correlations between the ventilator and the volumetric CO2 monitor for VD/VT were r = 0.97 and r(2) = 0.95 (P < .001), and the bias and precision were 0.01 ± 0.03.

Conclusions: PĒCO2, and therefore VD/VT, can be accurately calculated directly from the Dräger XL ventilator volumetric capnography measurements without use of a metabolic analyzer or volumetric CO2 monitor.

Keywords: dead space fraction; metabolic analyzer; ventilator; volumetric CO2 monitor; volumetric capnogram; volumetric capnography.

Publication types

  • Comparative Study

MeSH terms

  • Acute Lung Injury / diagnosis*
  • Acute Lung Injury / metabolism
  • Acute Lung Injury / physiopathology
  • Adult
  • Capnography* / methods
  • Capnography* / statistics & numerical data
  • Carbon Dioxide / analysis*
  • Dimensional Measurement Accuracy
  • Female
  • Humans
  • Male
  • Middle Aged
  • Monitoring, Physiologic / instrumentation
  • Monitoring, Physiologic / methods
  • Monitoring, Physiologic / statistics & numerical data
  • Observer Variation
  • Pulmonary Gas Exchange*
  • Respiratory Dead Space / physiology*
  • Respiratory Distress Syndrome / diagnosis*
  • Respiratory Distress Syndrome / metabolism
  • Respiratory Distress Syndrome / physiopathology
  • Statistics as Topic
  • Tidal Volume

Substances

  • Carbon Dioxide