Comparison of ThermoCool® Surround Flow catheter versus ThermoCool® catheter in achieving persistent electrical isolation of pulmonary veins: a pilot study

J Cardiovasc Electrophysiol. 2013 Mar;24(3):269-73. doi: 10.1111/jce.12031. Epub 2012 Dec 4.

Abstract

Introduction: Aim of this study was to compare efficacy and safety of the new ThermoCool Surround Flow® catheter (SFc) versus the ThermoCool® (TCc) in achieving persistent circumferential electrical isolation of the pulmonary veins (PVs) in patients with paroxysmal atrial fibrillation (AF).

Methods and results: This multicenter, randomized, controlled study enrolled patients suffering from paroxysmal AF. Randomization was run in a one-to-one fashion between radiofrequency ablation by TCc or SFc. Aim of PVs ablation was documentation of electrical isolation with exit/entrance block recorded on a circular catheter. Among the 106 enrolled patients, 52 (49.0%) were randomized to TCc and 54 (51.0%) to SFc. Total volume of infused saline solution during the procedure was lower in the SFc than in TCc group (752.7 ± 268.6 mL vs 1,165.9 ± 436.2 mL, P < 0.0001). Number of identified and isolated PVs was similar in the 2 groups. Number of PVs remaining isolated 30 minutes after ablation was higher in the SFc than in TCc group (95.2% vs 90.5%, P < 0.03), mainly driven by acute ablation result in the left PVs (96.1% vs 89.7%, P < 0.04). Complications were seldom and observed only in the TCc group (0% vs 3.84%, P < 0.03). At 6-month follow-up SFc patients reported a trend toward less AF recurrences compared to those in the TCc group (22.9% vs 27.0%, P = 0.69).

Conclusion: PV isolation by SFc lowered the rate of left PV early reconnections and reduced the volume of infused saline solution while maintaining the safety profile of AF ablation.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Aged
  • Atrial Fibrillation / diagnosis
  • Atrial Fibrillation / physiopathology
  • Atrial Fibrillation / surgery*
  • Cardiac Catheters*
  • Catheter Ablation / adverse effects
  • Catheter Ablation / instrumentation*
  • Chi-Square Distribution
  • Electrophysiologic Techniques, Cardiac
  • Equipment Design
  • Female
  • Humans
  • Italy
  • Male
  • Middle Aged
  • Pilot Projects
  • Predictive Value of Tests
  • Pulmonary Veins / physiopathology
  • Pulmonary Veins / surgery*
  • Recurrence
  • Sodium Chloride / administration & dosage
  • Therapeutic Irrigation / adverse effects
  • Therapeutic Irrigation / instrumentation*
  • Time Factors
  • Treatment Outcome

Substances

  • Sodium Chloride