Implementation of liquid chromatography/mass spectrometry into the clinical laboratory

Clin Chim Acta. 2013 May:420:4-10. doi: 10.1016/j.cca.2012.10.026. Epub 2012 Oct 22.

Abstract

Background: For certain clinical chemistry and toxicology analytes, liquid chromatography (LC) coupled with tandem mass spectrometry (MS/MS) offers significant advantages over traditional testing by immunoassay.

Methods: Published reports comparing the performance of immunoassays against LC-MS/MS were reviewed. The tested analytes include testosterone, estradiol, thyroid hormones, vitamin D, immunosuppressants, steroids for newborn screening programs, and clinical and forensic toxicology.

Results: While immunoassays are widely used in the clinical laboratory, the analytical sensitivity and specificity are inferior for many of the analytes tested in routine clinical laboratories. Moreover, LC-MS/MS can be multiplexed for high testing throughput and multiple analyte detection. The disadvantages of LC-MS/MS include the absence of Food and Drug Administration (FDA) and Conformité Européenne (CE) Mark approved tests, the high cost of analytical instrumentation, and the technical expertise needed to operate and maintain analyzers.

Conclusions: The implementation of LC-MS/MS will increase in the next few years as this technology continues to improve and the advantages become more well known.

Publication types

  • Review

MeSH terms

  • Chemistry, Clinical / methods*
  • Chromatography, Liquid*
  • Humans
  • Limit of Detection
  • Tandem Mass Spectrometry*