The resolution estimation of wedge and strip anodes

Rev Sci Instrum. 2012 Sep;83(9):093107. doi: 10.1063/1.4729117.

Abstract

A new resolution estimation method for wedge and strip anode based on the single photon imaging configuration is provided. The limiting resolution estimation equation is deduced theoretically according to the threshold principle. The relation between the charge cloud and the covered electrodes is discussed, and the equivalent diameter number is calculated. The resolution equation for the position deviation amplitude or FWHM is provided if noise exists. The relation between the position deviation amplitude and the total charge deviation amplitudes is discussed. The constancy of the position deviation amplitudes versus positions is provided. The results calculated from these equations are discussed. According to the equations, it is indicated that the spatial resolution is affected by the detection system configuration and noise. These conclusions may be useful for the designing and performance improvement of future photon imagers.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electrodes*
  • Photons*