Influence of different angles of reciprocation on the cyclic fatigue of nickel-titanium endodontic instruments

J Endod. 2012 Oct;38(10):1408-11. doi: 10.1016/j.joen.2012.05.019. Epub 2012 Jun 29.

Abstract

Introduction: The purpose of this study was to evaluate cyclic fatigue fracture resistance of engine-driven nickel-titanium (K3XF) instruments under reciprocating movement in various angles.

Methods: Fifty K3XF size 40 taper 0.06 nickel-titanium instruments were divided randomly into 5 groups of 10 each. All instruments were subjected to cyclic fatigue tests. Instruments in groups 1-4 were tested by using different reciprocating motions, whereas instruments of group 5 (control group) were used in continuous rotation. All instruments were rotated or reciprocated until fracture occurred. Time to fracture was recorded, and data were statistically analyzed by using one-way analysis of variance, followed by Tukey honestly significant difference test for comparison between different groups.

Results: All reciprocating groups (groups 1-4) showed a significant increase in time to failure when compared with group 5 (continuous rotation) (P < .05). Mean time was significantly higher in group 1, followed by group 2. No significant difference was found between groups 3 and 4 (P = .251). Increasing the clockwise angle of reciprocation and consequently increasing the angle of progression for each reciprocation cycle reduced the resistance to cyclic fatigue.

Conclusions: Movement kinematics (reciprocating movements in various angles) had a significant influence on the cyclic fatigue life of the tested nickel-titanium instruments.

MeSH terms

  • Biomechanical Phenomena
  • Dental Alloys*
  • Dental Instruments*
  • Dental Stress Analysis* / instrumentation
  • Equipment Design
  • Equipment Failure
  • Motion
  • Nickel
  • Random Allocation
  • Root Canal Preparation / instrumentation*
  • Rotation
  • Titanium
  • Torsion, Mechanical

Substances

  • Dental Alloys
  • titanium nickelide
  • Nickel
  • Titanium