Within-subject interlaboratory variability of QuantiFERON-TB gold in-tube tests

PLoS One. 2012;7(9):e43790. doi: 10.1371/journal.pone.0043790. Epub 2012 Sep 6.

Abstract

Background: The QuantiFERON®-TB Gold In-Tube test (QFT-GIT) is a viable alternative to the tuberculin skin test (TST) for detecting Mycobacterium tuberculosis infection. However, within-subject variability may limit test utility. To assess variability, we compared results from the same subjects when QFT-GIT enzyme-linked immunosorbent assays (ELISAs) were performed in different laboratories.

Methods: Subjects were recruited at two sites and blood was tested in three labs. Two labs used the same type of automated ELISA workstation, 8-point calibration curves, and electronic data transfer. The third lab used a different automated ELISA workstation, 4-point calibration curves, and manual data entry. Variability was assessed by interpretation agreement and comparison of interferon-γ (IFN-γ) measurements. Data for subjects with discordant interpretations or discrepancies in TB Response >0.05 IU/mL were verified or corrected, and variability was reassessed using a reconciled dataset.

Results: Ninety-seven subjects had results from three labs. Eleven (11.3%) had discordant interpretations and 72 (74.2%) had discrepancies >0.05 IU/mL using unreconciled results. After correction of manual data entry errors for 9 subjects, and exclusion of 6 subjects due to methodological errors, 7 (7.7%) subjects were discordant. Of these, 6 (85.7%) had all TB Responses within 0.25 IU/mL of the manufacturer's recommended cutoff. Non-uniform error of measurement was observed, with greater variation in higher IFN-γ measurements. Within-subject standard deviation for TB Response was as high as 0.16 IU/mL, and limits of agreement ranged from -0.46 to 0.43 IU/mL for subjects with mean TB Response within 0.25 IU/mL of the cutoff.

Conclusion: Greater interlaboratory variability was associated with manual data entry and higher IFN-γ measurements. Manual data entry should be avoided. Because variability in measuring TB Response may affect interpretation, especially near the cutoff, consideration should be given to developing a range of values near the cutoff to be interpreted as "borderline," rather than negative or positive.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Adult
  • Aged
  • Clinical Laboratory Techniques / methods*
  • Enzyme-Linked Immunosorbent Assay
  • Female
  • Humans
  • Interferon-gamma / blood
  • Male
  • Middle Aged
  • Observer Variation
  • Reagent Kits, Diagnostic*
  • Tuberculosis / blood
  • Tuberculosis / diagnosis*
  • Tuberculosis / epidemiology*
  • Young Adult

Substances

  • Reagent Kits, Diagnostic
  • Interferon-gamma

Grants and funding

The United States Air Force (USAF) funded this study as part of a larger project assessing reproducibility of the QuantiFERON-TB Gold In-Tube tests. The USAF and the Centers for Disease Control and Prevention (CDC) reviewed the study design, data collection methods, and analysis plans prior to approval. The USAF, U.S. Army, and CDC cleared the manuscript for publication according to established guidelines. No outside funders had a role in the study design, data collection, analysis, decision to publish, or preparation of the manuscript.