Capacitive immunoaffinity biosensor based on vertically paired ring-electrodes

Biosens Bioelectron. 2013 Feb 15;40(1):227-32. doi: 10.1016/j.bios.2012.07.028. Epub 2012 Aug 16.

Abstract

A capacitive biosensor was developed by using a vertically paired ring-electrode for the non-labeled immunoassay. The vertically paired ring-electrode was prepared by sequential deposition and etching processes. Two electrodes were layered on glass substrate by sputtering of gold layers with thicknesses of 50 nm and 100 nm, and a parylene-C film with a thickness of 550 nm was positioned between the electrode layers as a dielectric material by thermal deposition. The top layer was made by spin coating of SU-8. And then, the ring-electrodes were exposed at the wall of the layered structure by sequential etching processes. The fabricated electrodes were characterized by cyclic voltammetry of a well-known redox couple of 3,3',5,5'-tetramethylbenzidine. The non-labeled detection of antigen-antibody interaction was demonstrated by using anti-horseradish peroxidase (HRP) antibodies and C-reactive protein (CRP) as model analytes. When the model analytes were bound to the vertically paired ring-electrode, the impedance change was measured during the immunoassay steps, and the measured impedance was analyzed by using a model circuit of the ring-electrode, and the capacitance was estimated to be dependent on the adsorption of analytes between the ring-electrodes.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biosensing Techniques / instrumentation*
  • Conductometry / instrumentation*
  • Electric Capacitance
  • Electrodes*
  • Equipment Design
  • Equipment Failure Analysis
  • Immunoassay / instrumentation*
  • Reproducibility of Results
  • Sensitivity and Specificity