Measurement of teicoplanin by liquid chromatography-tandem mass spectrometry: development of a novel method

Ann Clin Biochem. 2012 Sep;49(Pt 5):475-81. doi: 10.1258/acb.2012.011257. Epub 2012 Aug 8.

Abstract

Background: Teicoplanin is an antibiotic used for the treatment of endocarditis, osteomyelitis, septic arthritis and methicillin-resistant Staphylococcus aureus. Teicoplanin is emerging as a suitable alternative antibiotic to vancomycin, where their trough serum levels are monitored by immunoassay routinely. This is the first report detailing the development of a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for measuring teicoplanin in patients' serum.

Methods: An Acquity™ UPLC (ultra-pressure liquid chromatography) tandem mass spectrometer was used to measure teicoplanin concentrations in samples from patients, quality assurance schemes and quality control preparations. Ristocetin was successfully implemented as a suitable internal standard. Ion suppression, linearity, stability, matrix effects, recovery, imprecision, lower limits of quantification and detection, interference and method comparison against immunoassay were all assessed.

Results: Teicoplanin and ristocetin had elution times of 1.39 and 1.24 min, respectively. Ion suppression was shown to be negligible, and linear calibration curves (0-200 μg/mL) were consistently reproduced to have r(2) values >0.99. Postextraction stability was achieved up to 20 h, while matrix effects were minimal coupled with sample recovery of >93%. The lower limit of quantification was 1 μg/mL, and 0.2 μg/mL was the lower limit of detection. Interference with other antibiotics was dependent on the combination of drugs present in patients' serum. A method comparison between immunoassay and LC-MS/MS suggested a negative bias for tandem mass spectrometry.

Conclusions: This novel method of teicoplanin determination by LC-MS/MS is proven to be a robust protocol that is consistent and reproducible. Clinicians searching for alternatives in therapeutic drug monitoring may have an additional option that is potentially more accurate and specific.

Publication types

  • Comparative Study
  • Validation Study

MeSH terms

  • Anti-Bacterial Agents / blood*
  • Blood Chemical Analysis / methods*
  • Chromatography, High Pressure Liquid*
  • Humans
  • Limit of Detection
  • Tandem Mass Spectrometry*
  • Teicoplanin / blood*

Substances

  • Anti-Bacterial Agents
  • Teicoplanin