[Evaluation of the Sysmex UF-1000i automated system for the diagnosis of urinary tract infection]

Enferm Infecc Microbiol Clin. 2013 Jan;31(1):29-31. doi: 10.1016/j.eimc.2012.05.017. Epub 2012 Aug 2.
[Article in Spanish]

Abstract

Objective: To evaluate the Sysmex UF-1000i system as a urine screening method for the diagnosis of urinary tract infection, and epithelial cells as a predictive value of contamination in woman of childbearing age.

Methods: A total of 1730 urine samples were processed using a urine culture as a reference.

Results: With 50 bacteria/μl as a cut-off point, the results were: sensitivity 91.3%, specificity 73.1%, negative predictive value 96.2%. For a specificity of 90% for epithelial cells, the results were: sensitivity 31.0%, positive predictive value 67.0%, negative predictive value 66.0%.

Conclusion: The evaluated system is fast and effective. Epithelial cells could be used to predict contamination.