High-Bandpass Filters in Electrocardiography: Source of Error in the Interpretation of the ST Segment

ISRN Cardiol. 2012:2012:706217. doi: 10.5402/2012/706217. Epub 2012 Jun 21.

Abstract

Introduction. Artifactual variations in the ST segment may lead to confusion with acute coronary syndromes. Objective. To evaluate how the technical characteristics of the recording mode may distort the ST segment. Material and Method. We made a series of electrocardiograms using different filter configurations in 45 asymptomatic patients. A spectral analysis of the electrocardiograms was made by discrete Fourier transforms, and an accurate recomposition of the ECG signal was obtained from the addition of successive harmonics. Digital high-pass filters of 0.05 and 0.5 Hz were used, and the resulting shapes were compared with the originals. Results. In 42 patients (93%) clinically significant alterations in ST segment level were detected. These changes were only seen in "real time mode" with high-pass filter of 0.5 Hz. Conclusions. Interpretation of the ST segment in "real time mode" should only be carried out using high-pass filters of 0.05 Hz.