Purpose: To assess the repeatability of common measurements with the Sirius Scheimpflug-Placido topographer and Lenstar LS900 optical low-coherence reflectometry (OLCR) biometer and the limits of agreement (LoA) between the devices.
Setting: Eye Hospital of Wenzhou Medical College, Wenzhou, China.
Design: Comparative evaluation of a diagnostic test or technology.
Methods: One randomly healthy eye of subjects was scanned 3 times with both devices. The parameters assessed were central corneal thickness (CCT), anterior chamber depth (ACD) from the corneal epithelium and from the endothelium, mean keratometry (K), and white-to-white (WTW) corneal diameter. The repeatability of scans was calculated using the within-subject standard deviation after 1-way analysis of variance was performed. The agreement between devices was assessed using the Bland-Altman LoA method, which equals the mean difference between devices ± 1.96 × standard deviation of the differences. The mean of 3 scans of each device was used to assess the LoA.
Results: Forty subjects were evaluated. The repeatability of the Scheimpflug-Placido topographer and OLCR biometer was 3.10 μm and 3.32 μm for CCT, 0.04 mm and 0.05 mm for WTW corneal diameter, and 0.17 D and 0.10 D for mean K, respectively. The repeatability for both devices was 0.02 mm for the ACD from the corneal epithelium and the ACD from the corneal endothelium. On Bland-Altman LoA analysis, all parameters were within clinically acceptable limits.
Conclusion: Both devices had excellent repeatability for all parameters assessed. Good LoAs were found between the 2 devices, indicating they can be used interchangeably for the parameters assessed.
Crown Copyright © 2012. Published by Elsevier Inc. All rights reserved.