Healing responses after bifurcation stenting with the dedicated TRYTON Side-Branch Stent™ in combination with XIENCE-V™ stents: a clinical, angiography, fractional flow reserve, and optical coherence tomography study: the PYTON (Prospective evaluation of the TRYTON Side-Branch Stent™ with an additional XIENCE-v™ everolimus-eluting stent in coronary bifurcation lesions) study

Catheter Cardiovasc Interv. 2013 Feb;81(3):E155-64. doi: 10.1002/ccd.24536. Epub 2012 Nov 14.

Abstract

Objectives: We evaluated healing responses with optical coherence tomography (OCT), and clinical and angiographic outcome after bifurcation stenting with the TRYTON Side-Branch Stent™.

Background: Dedicated bifurcation stents have been proposed as a potential alternative for treatment of true coronary bifurcation lesions.

Methods: We treated 20 consecutive patients with coronary bifurcation lesions and significant involvement of the side-branch (SB) with the TRYTON Stent and an additional XIENCE-V™ everolimus-eluting stent. At 9 months, we assessed the ratio of uncovered to total stent struts (RUTSS) with OCT, angiographic late luminal loss (LLL), and in-stent and in-segment restenosis. Clinical endpoints at 1 year included major adverse cardiac events (MACE) and their components [target lesion revascularization (TLR), myocardial infarction (MI), and cardiac death].

Results: LLL (N = 16) was 0.34 (0.17-0.46), 0.29 (0.24-0.48) and 0.57 (0.29-0.73) mm in the proximal main vessel (MV), distal MV and SB, respectively. In-bifurcation binary in-stent restenosis occurred in four patients (25%), in-segment restenosis in five (31.25%). The RUTSS (N = 13) was 4.0 ± 5.8, 0.7 ± 1.3, 0, and 2.5 ± 3.6% in the proximal MV, distal MV, SB, and polygon of confluence, respectively. At 1 year, MACE occurred in 5 (25%) [4 TLR (20%), 3 MI (15%)].

Conclusion: The homogeneous stent strut coverage and the low LLL in the MV reflect proper healing characteristics of the TRYTON Stent in combination with the XIENCE-V™ stent. However, proximal MV edge and ostial SB restenoses together with overall clinical outcomes do not fulfill expectations of a dedicated bifurcation stent. © 2012 Wiley Periodicals, Inc.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Antineoplastic Agents
  • Coronary Angiography / methods*
  • Coronary Circulation / physiology*
  • Coronary Stenosis / diagnosis*
  • Coronary Stenosis / physiopathology
  • Coronary Stenosis / surgery
  • Coronary Vessels / pathology
  • Coronary Vessels / physiopathology
  • Coronary Vessels / surgery
  • Drug-Eluting Stents*
  • Everolimus
  • Female
  • Humans
  • Immunosuppressive Agents / pharmacology
  • Male
  • Myocardial Revascularization / methods*
  • Prospective Studies
  • Prosthesis Design
  • Regional Blood Flow
  • Sirolimus / analogs & derivatives*
  • Sirolimus / pharmacology
  • Tomography, Optical Coherence / methods*
  • Treatment Outcome

Substances

  • Antineoplastic Agents
  • Immunosuppressive Agents
  • Everolimus
  • Sirolimus