Pilot Evaluation of RT-PCR/Electrospray Ionization Mass Spectrometry (PLEX-ID/Flu assay) on Influenza-Positive Specimens

Open Virol J. 2012:6:64-7. doi: 10.2174/1874357901206010064. Epub 2012 May 9.

Abstract

The PLEX-ID/Flu assay has been recently developed to enable the detection and typing of influenza viruses based on the RT-PCR/electrospray ionization mass spectrometry technology.This novel assay was evaluated for typing performance on 201 positive influenza A or B nasopharyngeal swab specimens (NPS) detected by real-time RT-PCR during the 2010-2011 season. The PLEX-ID/Flu assay detected and characterized 91.3% and 95.3% of all influenza A and B samples, respectively. All non-typeable influenza A and B specimens by the assay showed low viral loads with threshold cycle values ≥ 33. Taken together, and although our results need to be confirmed by further prospective studies, the PLEX-ID/Flu assay detected positively and gave a typing result for 93% of all NPS detected positively by real-time RT-PCR, thus suggesting a potential role for influenza virus surveillance among other techniques.

Keywords: Influenza surveillance; PLEX-ID/Flu assay.; influenza typing.