Microwave ablation treatment of liver cancer with a 2,450-MHz cooled-shaft antenna: pilot study on safety and efficacy

Asian Pac J Cancer Prev. 2012;13(2):737-42. doi: 10.7314/apjcp.2012.13.2.737.

Abstract

To evaluate efficacy of microwave ablation in a primary clinical study, sixty patients (44 men, 16 women; mean age 53 years) with 96, 1-8 cm (mean 3.20 ± 0.17 cm) liver cancers were treated with 2,450-MHz internally cooled-shaft antenna. Complete ablation (CA) and local tumor progression (LTP) rates as well as complications were determined. CA rates in small (<3.0 cm), intermediate (3.1-5.0 cm) and large (5.1-8.0 cm) liver cancers were 96.4% (54/56), 92.3% (24/26) and 78.6% (11/14), respectively. During a mean follow-up period of 17.17 ∓ 6.52 months, LTP occurred in five (5.21%) treated cases. There was no significant difference in the CA and LTP rates between the HCC and liver metastasis patient subgroups (P<0.05). Microwave ablation provides a reliable, efficient, and safe technique to perform hepatic tumor ablation.

MeSH terms

  • Adult
  • Aged
  • Carcinoma, Hepatocellular / diagnostic imaging
  • Carcinoma, Hepatocellular / secondary
  • Carcinoma, Hepatocellular / surgery*
  • Catheter Ablation / instrumentation*
  • Cold Temperature
  • Electrocoagulation
  • Female
  • Follow-Up Studies
  • Humans
  • Liver Neoplasms / diagnostic imaging
  • Liver Neoplasms / pathology
  • Liver Neoplasms / surgery*
  • Male
  • Microwaves / therapeutic use*
  • Middle Aged
  • Pilot Projects
  • Prognosis
  • Tomography, X-Ray Computed
  • Young Adult