Successful retrieval of a fractured and entrapped 0.035-inch terumo wire in the femoral artery using biopsy forceps

Korean Circ J. 2012 Mar;42(3):201-4. doi: 10.4070/kcj.2012.42.3.201. Epub 2012 Mar 26.

Abstract

A 0.035-inch guide wire fracture and entrapment in a peripheral artery is a very rare complication, but when it does occur it may lead to life-threatening complications, such as perforation, thrombus formation, embolization, and subsequent limb ischemia. We describe our experience of successfully retrieving a fractured 0.035-inch Terumo guide wire in the external iliac artery using a biopsy forcep.

Keywords: Complications; Percutaneous transluminal coronary angioplasty.