Practical considerations for in vivo MRI with higher dimensional spatial encoding

MAGMA. 2012 Dec;25(6):419-31. doi: 10.1007/s10334-012-0314-y. Epub 2012 Apr 7.

Abstract

Object: This work seeks to examine practical aspects of in vivo imaging when spatial encoding is performed with three or more encoding channels for a 2D image.

Materials and methods: The recently developed 4-Dimensional Radial In/Out (4D-RIO) trajectory is compared in simulations to an alternative higher-order encoding scheme referred to as O-space imaging. Direct comparison of local k-space representations leads to the proposal of a modification to the O-space imaging trajectory based on a scheme of prephasing to improve the reconstructed image quality. Data were collected using a 4D-RIO acquisition in vivo in the human brain and several image reconstructions were compared, exploiting the property that the dense encoding matrix, after a 1D or 2D Fourier transform, can be approximated by a sparse matrix by discarding entries below a chosen magnitude.

Results: The proposed prephasing scheme for the O-space trajectory shows a marked improvement in quality in the simulated image reconstruction. In experiments, 4D-RIO data acquired in vivo in the human brain can be reconstructed to a reasonable quality using only 5 % of the encoding matrix--massively reducing computer memory requirements for a practical reconstruction.

Conclusion: Trajectory design and reconstruction techniques such as these may prove especially useful when extending generalized higher-order encoding methods to 3D images.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Artifacts
  • Brain / pathology
  • Calibration
  • Computer Simulation
  • Computers
  • Fourier Analysis
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Imaging, Three-Dimensional / methods*
  • Magnetic Resonance Imaging / methods*
  • Models, Statistical
  • Software