Influence of the thorium decay series on the background of high-resolution gamma-ray spectrometers

Appl Radiat Isot. 2012 Jun;70(6):1005-9. doi: 10.1016/j.apradiso.2012.03.014. Epub 2012 Mar 17.

Abstract

The background induced by the members of the thorium decay sequence in six high-resolution, gamma-ray spectrometers was analyzed. For the analysis, the count rates in the peaks of the background spectra, normalized to the unit of emission probability and detection probability, were used. The energy dependence of these normalized count rates carries information about the location of the sources of contamination. The contributions of the detector contamination, the contamination of the shielding material and the radiation penetrating the shield were calculated. The comparison of these contributions among the spectrometers pointed to the weaknesses of some shields, making such a comparison a useful tool for assessing the effectiveness of the shields.

Publication types

  • Evaluation Study

MeSH terms

  • Background Radiation*
  • Equipment Contamination*
  • Equipment Design
  • Equipment Failure Analysis
  • Radiation Protection / instrumentation*
  • Spectrometry, Gamma*
  • Thorium / analysis*

Substances

  • Thorium