Half-life determination of 88Kr and 138Xe

Appl Radiat Isot. 2012 Sep;70(9):1927-30. doi: 10.1016/j.apradiso.2012.02.054. Epub 2012 Feb 25.

Abstract

The half-lives of (88)Kr and (138)Xe have been determined with two high purity germanium (HPGe) detectors by the position relay method. Data have been recorded at regular time intervals during measurements covering more than nine times the respective half-lives. This article describes in detail the principle and process of the position relay method with two HPGe detectors. The determined half-lives of (88)Kr and (138)Xe are 2.804±0.015 h and 14.18±0.10 min, respectively.

MeSH terms

  • China
  • Half-Life
  • Internationality
  • Krypton Radioisotopes / analysis*
  • Krypton Radioisotopes / chemistry*
  • Radiation Dosage
  • Radiometry / instrumentation*
  • Radiometry / standards*
  • Reference Standards
  • Reference Values
  • Xenon Radioisotopes / analysis*
  • Xenon Radioisotopes / chemistry*

Substances

  • Krypton Radioisotopes
  • Xenon Radioisotopes